MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problems Insufficient Cooling (1130); Temperature Problem (3022); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
 
Event Description
It was reported that the patient was not cooling on the arctic sun device. The device displayed alert 113. The patient temperature was 35. 5c, target was 33c, and water temperature was 26c. The nurse did not see the water get lower than that. Ms&s explained that the device will need to be taken out of service and sent to the biomed, labeled alert 113 and broken cooler. Per follow up, (b)(6) stated the patient was able to complete therapy on the second device with no further issues and that the first device would be sent to biomed. Biomed was still in the process of looking at the device and planned to repair it at the facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key9596012
MDR Text Key180401814
Report Number1018233-2020-00348
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/25/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-