• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH NON CONNECT 4TH ED DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH NON CONNECT 4TH ED DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66802134
Device Problem Electrical Power Problem (2925)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2019
Event Type  malfunction  
Event Description
It was reported that the device presented a battery issue: the console does not hold the charge and a "burning smell" was also reported by the medical team, it was confirmed that the device was not overheating at the moment of the malfunction.The device is available for analysis.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, has been returned for servicing only, not fault was found with the device, as such we are unable to establish a relationship between the event report and the device.A review of the device history showed that this unit passed all acceptance criteria and was compliant upon release for distribution.There were no indications to suggest the device did not meet product specifications nor did it allege any product deficiencies.Complaint history has been reviewed with similar instances recorded in the previous four years, which are under constant review monitoring for adverse trends.Battery failure can occur due to certain situations: 1.The lithium ion re-chargeable battery, contained within the device is subject to a limited number of charge cycles, battery failure can occur when it has reached its life expectancy.2.The renasys device is also fitted with a safety design feature, which activates when the battery detects that it has achieved a certain temperature, charging will pause.The instructions for use offer detailed guidance on steps taking with regard to charging the device.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH NON CONNECT 4TH ED DEVICE
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull
UK 
MDR Report Key9596147
MDR Text Key175229577
Report Number8043484-2020-00034
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
PMA/PMN Number
K152163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66802134
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2020
Date Manufacturer Received06/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-