• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN TOUCH SCRO ZERO ANG 22CM; INFLATABLE PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN TOUCH SCRO ZERO ANG 22CM; INFLATABLE PENILE PROSTHESIS Back to Search Results
Model Number ES29221022
Device Problems Break (1069); Material Twisted/Bent (2981); Material Split, Cut or Torn (4008)
Patient Problem No Information (3190)
Event Date 12/19/2019
Event Type  Injury  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa review.No trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to the available information, kinking/split in left tubing near pump were reported.The device was explanted and replaced with another device.
 
Manufacturer Narrative
This follow-up mdr is created to document the evaluation of the returned device.A titan touch pump and two cylinders were received for evaluation.Examination and testing of the returned components revealed a separation within abrasion in the shorter exhaust tube of the pump.Testing revealed this to be a site of leakage.Partial separations within abrasion were noted in the shorter length pump exhaust tube and pump inlet tube.Testing revealed these to not be sites of leakage.Based on examination of the returned product, it was concluded that the abrasion marks noted on all pump tubing matched in such a way to conclude that they had overlapped and abraded against one another while in-vivo.This positioning, in combination with device usage over time, could most likely contribute to sufficient stress(s) to separate the exhaust tubing of the shorter length pump exhaust tubing.A separation of this type would then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN TOUCH SCRO ZERO ANG 22CM
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebaek, 3050
DA  3050
MDR Report Key9599392
MDR Text Key175444445
Report Number2125050-2020-00052
Device Sequence Number1
Product Code FHW
Combination Product (y/n)N
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberES29221022
Device Catalogue NumberES2922
Device Lot Number5164896
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2020
Date Manufacturer Received12/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age58 YR
-
-