|
Model Number 2420-0007 |
Device Problems
Fluid Leak (1250); Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
No product will be returned per customer.
The customer complaint could not be confirmed because the product was discarded and not returned for failure investigation.
The root cause of this failure was not identified.
Patient information requested but not provided, however the customer stated that the patient was an adult patient.
|
|
Event Description
|
It was reported from cardio thoracic icu that the nurse initiated a primary continuous fentanyl infusion at 0828 per md.
Approximately around 1000, the nurse noted fluid on the floor.
Upon further investigation, a hole located approximately 1 inch below the pump was noted in the tubing and the bag appeared to have less fluid than expected after only being hung 1.
5 hours.
The tubing set was immediately replaced and no further issues were reported.
It was reported that there was no impact to the adult patient from the event.
|
|
Search Alerts/Recalls
|
|
|