It was reported that the during an acl and shoulder arthroscopy procedures, the dyonics control unit kept stopping several times.The procedure was completed with significant delay greater than 60 minutes using the same device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H3, h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of intermittent pump function could not be reproduced.Product passed functional testing per factory test with no faults or errors.Product passed functional testing during 2 hour burn-in on wet station utilizing low and high pressure and flow settings.Raw and zero transducer readings were normal and well within specs during all functional tests.At no time during functional testing did pump stop intermittently.All functions perform as expected.Pressure and flow readings were normal throughout burn-in on wet station.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures.No containment or corrective actions are recommended at this time.
|