• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR LINERS Back to Search Results
Catalog Number UNK HIP ACETABULAR LINER
Device Problems Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Patient Problems Foreign Body Reaction (1868); Pain (1994); No Code Available (3191)
Event Date 04/08/2013
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Pinnacle claim letter and medical records received.Claim letter alleges metallosis and elevated chromium level.After review of medical records, the patient was revised to address pain.X-ray showed evidence of wear.There was significant metallosis secondary to the ball rubbing on the cup after the polyethylene had worn through.The polyethylene was dislodged from the cup.There was a small amount of burnishing on the inside of the cup.The cup was removed as well.Doi: (b)(6) 2011; dor: (b)(6) 2013; (left hip).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP ACETABULAR LINERS
Type of Device
HIP ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380-0988
6103142063
MDR Report Key9611787
MDR Text Key176255716
Report Number1818910-2020-02401
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR LINER
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-