• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number UNK-P-IPP
Device Problems Break (1069); Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2019
Event Type  Injury  
Manufacturer Narrative
The complaint component was returned and analyzed, and the reported allegation was confirmed via product analysis.The cylinders returned cut in half.Product analysis identified a fluid leak in the body of cylinder 2 along with holes in the outer tubes of both cylinders.The fluid leak was attributed to input tube wear and fatigue.The holes in the outer tube were also attributed to input tube wear.This damage would affect the functionality of the device and would therefore be the most probably cause for the reported events.The event cannot be reproduced or substantiated; therefore, no escalation to ncep, capa or scar is required.
 
Event Description
It was reported that the patient underwent a removal of an inflatable penile prosthesis (ipp) device which was implanted back in 1995.The removal for the device was due to cylinder rupture on both sides.So both cylinders had ruptured and all the fluid had leaked out so the device failed.The doctor removed the device and replaced it with a malleable tactra penile prosthesis device.The reason for this revision was device malfunction.The patient is doing well.No more information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
alyson harris
10700 bren road w
minnetonka, MN 55343
4089353452
MDR Report Key9613161
MDR Text Key175852256
Report Number2183959-2020-00183
Device Sequence Number1
Product Code FHW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUNK-P-IPP
Device Catalogue NumberUNK-P-IPP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
-
-