The complaint component was returned and analyzed, and the reported allegation was confirmed via product analysis.The cylinders returned cut in half.Product analysis identified a fluid leak in the body of cylinder 2 along with holes in the outer tubes of both cylinders.The fluid leak was attributed to input tube wear and fatigue.The holes in the outer tube were also attributed to input tube wear.This damage would affect the functionality of the device and would therefore be the most probably cause for the reported events.The event cannot be reproduced or substantiated; therefore, no escalation to ncep, capa or scar is required.
|