MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED2-475-30
Device Problems Break (1069); Retraction Problem (1536)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/16/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that the distal tip of the pipeline pushwire broke off inside the patient.The patient was undergoing surgery for treatment of a saccular, unruptured, right ic aneurysm with a max diameter of 12 mm and a 4.7 5x3.5 mm landing zone.It was noted the patient's vessel tortuosity was moderate.It was reported that after deploying the pipeline, the pushwire was attempted to be retracted into the phenom27 microcatheter, but r esistance was felt.The phenom was retrieved without being retracted.An angiogram confirmed the distal tip of the pushwire was left inside the body.A 4mm gooseneck snare was used to successfully recover the pushwire tip.There was no resistance or difficulty in the operation and the stent was successfully placed.Postoperative blood flow angiography showed no problems.It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event.Ancillary devices include an envoy6fr guide catheter, phenom 27 microcatheter, and a chikai014 guidewire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Type of Device
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
katcha taylor
9775 toledo way
irvine, CA 92618
9496801345
MDR Report Key9617722
MDR Text Key186774718
Report Number2029214-2020-00064
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPED2-475-30
Device Catalogue NumberPED2-475-30
Device Lot NumberA822946
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-