• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TSH ASSAY RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TSH ASSAY RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number TSH
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/17/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6). All tsh results from the roche analyzers and the analyzer from abbott were below the normal reference ranges for the respective assays. Calibration and qc at the investigation site were ok. A general reagent issue can be excluded. Assays from different manufacturers, in this case abbott, can generate different results. This relates to the overall setups of the assays, the antibodies used, differences in reference materials and the standardization methodology used. The investigation did not identify a product problem.
 
Event Description
The initial reporter questioned thyroid results for 1 patient sample tested on a cobas e801 module. The sample was submitted for investigation where discrepant results were identified for elecsys tsh (tsh) and elecsys ft4 iii (ft4 iii) between the customer's e801 module, an e801 module used at the investigation site and the abbott architect method. This medwatch will cover tsh. Refer to medwatch with a1 patient identifier (b)(6) for information on the ft4 iii results. Refer to attached data for the patient results. No questionable results were reported outside of the laboratory. The customer's e801 module serial number was not provided. The e801 module serial number used at the investigation site was (b)(4). The tsh reagent lot number used at the investigation site was 416233 with an expiration date of oct-2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameELECSYS TSH ASSAY
Type of DeviceRADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM 68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key9621736
MDR Text Key214801102
Report Number1823260-2020-00188
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/23/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberTSH
Device Catalogue Number07028091190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-