Medtronic received a report that the pipeline pushwire broke off in the patient.
The patient was undergoing surgery for treatment of a saccular, unruptured, cavernous right ica aneurysm with a max diameter of 9mm, a 5mm neck diameter, and a landing zone of 4.
8x5mm.
It was noted the patient's vessel tortuosity was moderate.
It was reported that the pipeline was deployed in the m1, and when the physician started to drag the device back, the pushwire broke off from the distal tip proximal to the re-sheathing marker.
A snare was used to try and retrieve the wire, but it didn't work.
Four atlas stents were placed to cover the wire from the pcom (posterior communicating artery) and into the cavernous section past the pipeline but before the aneurysm.
It was indicated that all devices were prepared as per the instructions for use (ifu), and it was stated the patient was doing well.
Ancillary devices include a 6fr shuttle sheath, phenom 027 microcatheter, phenom plus guide catheter, and a synchro 2 guidewire.
|