MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Blower (738); Defective Component (2292); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Malfunction  
Manufacturer Narrative

Date of event: (b)(6) 2020. Date of report: 01/23/2020.

 
Event Description

The biomedical engineer (bme) identified during preventive maintenance there was an issue with a failed blower, back up battery, and flow sensor assembly. The bme confirmed the reported failures. The bme replaced the blower, back up battery, and flow sensors to correct the identified failures. The bme has reported the unit has been returned to service. The device was not in use at the time of the reported issue, and no patient or user harm was reported.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameV60 VENTILATOR
Type of DeviceVENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key9624970
MDR Text Key183950466
Report Number2031642-2020-00304
Device Sequence Number1
Product Code MNT
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation
Type of Report Initial
Report Date 12/27/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received01/23/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator NO INFORMATION
Device MODEL NumberV60
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/27/2019
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured12/16/2009
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-