U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-45-FEM
Device Problems Structural Problem (2506); Device Tipped Over (2589); Insufficient Information (3190)
Patient Problems Chest Pain (1776); Internal Organ Perforation (1987); Perforation of Vessels (2135); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
It is alleged that the patient received a gunther tulip on (b)(6) 2005.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Event Description
Patient allegedly received an implant on (b)(6) 2005 via the femoral vein due to deep vein thrombosis (dvt).Patient is alleging tilt, vena cava perforation, organ perforation (mesenteric and gastrointestinal tract).Patient further alleges small bowel perforation, chest pain, anxiety, mental anguish, stress, worry.On (b)(6) 2017, report from computerized tomography (ct): "there is an ivc filter in place which is oriented para level to the orientation para level of the ivc.The tip is located at the level of the single left renal vein and just below the origin of the right renal vein.The four prongs penetrate the wall of the ivc.The 2 anterior prongs at the 10:00 and 2:00 position of but the second/third portion of the duodenum with the leftward most prong at least tenting and possibly penetrating the duodenum.The posterior prongs do not extend into the adjacent organs.The ivc is normal caliber with no evidence of stenosis.".
 
Manufacturer Narrative
Additional information: investigation the following allegations have been investigated: vena cava (vc)/organ/mesenteric/gi/small bowel perforation, tilt, pain, anxiety, mental anguish, stress, worry.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain, anxiety, mental anguish, stress, and worry are directly related to the filter and unable to identify a corresponding failure mode at this point in time.A total of 10 devices were manufactured in the reported lot.To date, no other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9627226
MDR Text Key176253741
Report Number1820334-2020-00201
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2005
Device Catalogue NumberIGTCFS-45-FEM
Device Lot Number1504046
Was Device Available for Evaluation? No
Date Manufacturer Received09/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
-
-