PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Patient Involvement (2645)
Event Date 01/02/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to physio-control for evaluation.The cause of the reported issue could not be determined.Device not evaluated by manufacturer.
 
Event Description
The customer contacted physio-control to report that their device logged a critical event code and is not passing the self-test.The event code is indicative of a device issue that could result in a partial loss of defibrillator output energy due to a loss of the negative or positive portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key9634731
MDR Text Key191105991
Report Number0003015876-2020-00121
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873813744
UDI-Public00883873813744
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number99402-000016
Was Device Available for Evaluation? No
Date Manufacturer Received01/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/26/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-