WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number IGTCFS-65-1-FEM-CELECT-PT |
Device Problems
Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Vessel Or Plaque, Device Embedded In (1204); Perforation of Vessels (2135); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
Manufacturers ref# (b)(4).Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2015.It is alleged that the [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.".
|
|
Event Description
|
Patient allegedly received an implant on 22jul2015 via the right common femoral vein due to being status post bleed.Patient is alleging device tilt, vena cava perforation and embedment.Per a 18jan2018 ct2 (computed tomography 2): "impressions: the filter is tilted posteriorly.Its proximal cone lies on the wall of the ivc possible embedded in it.The filter legs have penetrated through the wall of the ivc into the pericaval/mesenteric fat".
|
|
Manufacturer Narrative
|
H6 (device code): appropriate term/code not available (3191) for alleged device perforation.H6 (device code): appropriate term/code not available (3191) for alleged device tilt.Investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava perforation, embedment, tilt.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.20 devices in lot.To date, no other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
Conclusion code(s): appropriate term/code not available (4316) was selected because the previous investigation remains unchanged by the additional information.Investigation: investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.Cook will reopen the investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.20 devices in lot.To date, no other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
No additional information provided at this time.
|
|
Manufacturer Narrative
|
Additional information: investigation.Investigation is reopened due to additional information provided.The following allegations have been investigated: pain.The reported allegations have been further investigated based on the information provided to date.Unknown if the reported pain is directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient further alleges pain.
|
|