• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problem Entrapment of Device (1212)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/16/2019
Event Type  malfunction  
Manufacturer Narrative
A sample product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There are no other complaints in the lot.The manufacturer internal reference number is: (b)(4).
 
Event Description
A facility representative reported that during an intraocular lens (iol) implant procedure, the lens became stuck in the preloaded injector.There was patient contact, but patient harm.Additional information was requested.
 
Manufacturer Narrative
The device with the lens was returned loose in the carton.The plunger lock and lens stop have been removed.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger was retracted into the barrel of the device, prior to the loading entry area.The lens was located within the loading area.The trailing haptic is folded onto the optic.The leading haptic is extended into the nozzle area.The nozzle was cleaned for further evaluation.The lens was removed during cleaning.Top coat dye stain testing was conducted with acceptable results.All product and batch history records are quality reviewed prior to product release.A qualified viscoelastic was used.The root cause cannot be determined for the complaint of ¿lens stuck in system--with patient contact¿.The lens was slightly advanced/stuck within the loading area upon return.The trailing haptic is folded onto the optic.The positon of the lens in the returned device would suggest that the lens did not advance beyond the loading area.The plunger was retracted into the barrel prior to the loading area.Additional information was provided in d.10., h.3., h.6.And h.10.Corrected information provided in h.6.Result code.In prior mdr 3220-mutagenic problem was reported in error.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key9639007
MDR Text Key181379678
Report Number1119421-2020-00123
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Model NumberAU00T0
Device Lot Number12579492
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2020
Date Manufacturer Received03/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNSPECIFIED PROVISC OVD.
-
-