MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3562
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Twiddlers Syndrome (2114); No Known Impact Or Consequence To Patient (2692)
Event Date 01/15/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer number: 2017865-2020-00826, 2017865-2020-00827, 2017865-2020-00829.It was reported that the patient presented in clinic.It was noted that remote transmissions showed low sensing on the atrial lead and loss of capture on the right and left ventricular leads.Dislodgement of all leads into the superior vena cava, twisted into the pocket was observed.The physician believed any anomalies were related to the patient, mechanical and as a result of twiddlers syndrome.The entire system was explanted.The patient was stable.
 
Manufacturer Narrative
(b)(4).Product problem should not be a selection as the pacemaker was explanted due to twiddler's.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9645574
MDR Text Key176933939
Report Number2017865-2020-00825
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000081386
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/15/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET; TENDRIL STS; TENDRIL STS
Patient Outcome(s) Required Intervention;
-
-