Information was received from a foreign healthcare professional (hcp) via a manufacturer representative regarding a patient receiving intrathecal clonidine 1,000.0 mcg/ml at 74.9 mcg/day via an implantable pump.The indication for use was not reported.It was reported that when the customer interrogated the pump on (b)(6) 2020 at 15:15, the service codes 101 (a pump reset has occurred.Infusion is set to minimum rate.), 87 (potential loss of therapy : the pump is in safe mode.Therapy is running in minimum rate.), and 116 ( potential underdose : due to pump memory error, myptm may not accurately access the myptm settings.The issue will be resolved at update.) appeared on the tablet screen.The customer then closed the session.The patient did not show any withdrawal symptoms at the interrogation.No alarm was reported to go off.After contacting the device manufacturer, the device manufacturer technical services department recommended that they clear the warnings and program the pump according to the therapeutic needs of the patient.The pump was therefore re-interrogated and reprogrammed on (b)(6) 2020 at ¿15:20¿ [logs indicated that the pump was actually re-programmed around 15:32 to 15:33].The patient was observed for some time, but no sign of underdose or overdose was reported.Logs showed: critical alarm - pump defaulted to minimum rate (07) on (b)(6) 2020 at 15:16 critical alarm - pump reset due to firmware error (2c) on (b)(6) 2020 at 15:16 critical alarm - pump reset (01) on (b)(6) 2020 at 15:16 critical alarm - pump defaulted to minimum rate (07) on (b)(6) 2020 at 15:32 pump motor stall recovery (1a) on (b)(6) 2020 at 15:33.No further complications were reported.
|