ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN001112
Device Problem Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/10/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported the intra-aortic balloon pump (iabp) touchscreen froze up during use.The staff exchanged for another iabp after trying screen calibration, cursor movement and power recycling the iabp.The patient had remained stable and supported on the iabp with no other noted alarms or issues.Patient was successfully supported on the second iabp without issue.There was no report of patient complications, serious injury or death.
 
Manufacturer Narrative
Qn#(b)(4).No iabp part or recorder strip has been returned to teleflex chelmsford for investigation.The reported complaint of iabp touch screen unresponsive is not able to be confirmed.Per the event details, the pump was powered on after being powered off for more than 5 minutes, and the issue did not reoccur.The root cause of the complaint is undetermined.Device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
 
Event Description
It was reported the intra-aortic balloon pump (iabp) touchscreen froze up during use.The staff exchanged for another iabp after trying screen calibration, cursor movement and power recycling the iabp.The patient had remained stable and supported on the iabp with no other noted alarms or issues.Patient was successfully supported on the second iabp without issue.There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key9649379
MDR Text Key184648048
Report Number3010532612-2020-00010
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902084966
UDI-Public30801902084966
Combination Product (y/n)N
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN001112
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? No
Date Manufacturer Received02/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-