MAQUET CARDIOPULMONARY GMBH HEATER COOLER UNIT; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701034642 - HCU 30
Device Problems Insufficient Heating (1287); Temperature Problem (3022); Radiation Output Problem (4026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/28/2020
Event Type  malfunction  
Manufacturer Narrative
A follow up medwatch will be submitted when additional information becomes available.
 
Event Description
(b)(4).It was reported that there was no circulation of cool water during operation.To complete the operation, they had to put hot water into the water tank to melt the ice block.No health damage to the patient.
 
Event Description
Complaint (b)(4).
 
Manufacturer Narrative
The reported failure "no circulation of cool water" could be confirmed.According to the communication in getinge complaint system dated 2020-02-19 the getinge field service technician (fst) informed that the ice sensor of the hcu 30 device was replaced.The replaced ice sensor was requested for further investigation in life cycle engineering (lce) with rma #(b)(4) on 2020-03-13.Part was received on 2020-04-09.Lce performed the investigation according to report #(b)(4) on 2020-04-20: the investigated ice sensor was destroyed through corrosion.Electrical resistances for two out of three ntcs (negative temperature coefficient thermistor) showed readings significantly off spec.The most probably root caused was corrosion by water and cleaning agents leaking into the sensor.The reported failure "no circulation of cool water" occurred during patient treatment.The hcu 30 which was used, was responsible for this complaint.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopumonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER COOLER UNIT
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key9652827
MDR Text Key190536419
Report Number8010762-2020-00042
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K031544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 04/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701034642 - HCU 30
Device Catalogue Number701034642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2020
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-