MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID INT TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID INT TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Failure to Power Up (1476); Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was reviewed and no non-conformance's related to the reported event were noted. The getinge service territory manager (stm) replaced the power management board and the issue was resolved. The iabp functioned normally. The initial reporter named is the getinge designee who has different contact details from that of the event site: (b)(6).
 
Event Description
It was reported that during an inspection of the cardiosave intra-aortic balloon pump (iabp) prior to delivery to the hospital, the power supply did not start. Since the event occurred during inspection, there was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID INT TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key9652828
MDR Text Key191554895
Report Number2249723-2020-00155
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 04/16/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-65
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/16/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-