Additional information was added: two (2) devices were received for evaluation.
A visual inspection was performed, and it was noted that one of the returned samples was a complete set with no abnormalities observed.
The second returned sample was an incomplete set with the tubing cut approximately 50 cm away from the drip chamber.
During the investigation it was established that the second returned sample was cut at a cross-cutter station during the assembly process.
The tubing was dislodged from the blister pocket and cut at the cross-cutter station during assembly.
The reported condition was not verified for sample one; however, was verified for the second returned sample.
The cause of the reported condition was a manufacturing issue.
Should additional relevant information become available, a supplemental report will be submitted.
|