Catalog Number 31-473601 |
Device Problem
Fracture (1260)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the during hip surgery the universal inserter/extractor impaction plate fractured while trying to remove the stem.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Reported event was confirmed by visual inspection of reported device.Visual inspection identified the weld on the impact plate had fractured causing the plate to spin freely.The plate cannot be pulled off of the device without damaging the remaining weld.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|