• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122538
Device Problems Noise, Audible (3273); Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Discomfort (2330); Injury (2348)
Event Date 06/03/2014
Event Type  Injury  
Event Description
It was reported a left hip revision surgery due to pain and discomfort as well as a rubbing sensation in left hip that created an audible squeaking noise and black tissue.Head removed.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision, the hemi head and modular sleeve were removed.The bhr cup and stem remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup, hemi head and modular sleeve was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the hemi head.Similar complaints have been identified for the sleeve.However, as the device is no longer sold, no action is to be taken.Similar complaints have been identified for the bhr cup.However, as bhr systems of this size are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.Based on the information provided the reported squeaking, elevated cocr levels and intraoperative findings of fluid and black discolored tissue may be consistent with findings associated with metal debris and trunnionosis.Without supporting postprimary and pre-revision radiographic images, and/or the analysis of the explanted components, the source of the reported findings cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the revision and expected post-operative healing/pain cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
New information: g4, d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key9659472
MDR Text Key177400241
Report Number3005975929-2020-00031
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date06/30/2012
Device Catalogue Number74122538
Device Lot Number07DW11422
Was Device Available for Evaluation? No
Date Manufacturer Received08/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
71356004 ANTH SO PO SIZE 4 07DM03420A; 74120148 ACETLR CUP HAP 48MM W/ IM 75040; 74222200 MOD SLEEVE PLUS 0MM 12/14 07FW11497
Patient Outcome(s) Hospitalization; Required Intervention;
-
-