It was reported that left hip revision surgery was performed.During the revision, the hemi head and modular sleeve were removed.The bhr cup and stem remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup, hemi head and modular sleeve was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the hemi head.Similar complaints have been identified for the sleeve.However, as the device is no longer sold, no action is to be taken.Similar complaints have been identified for the bhr cup.However, as bhr systems of this size are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.Based on the information provided the reported squeaking, elevated cocr levels and intraoperative findings of fluid and black discolored tissue may be consistent with findings associated with metal debris and trunnionosis.Without supporting postprimary and pre-revision radiographic images, and/or the analysis of the explanted components, the source of the reported findings cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the revision and expected post-operative healing/pain cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|