Catalog Number 121887458
Device Problems Naturally Worn (2988); Noise, Audible (3273)
Patient Problems Joint Disorder (2373); No Code Available (3191)
Event Date 11/23/2011
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4). Investigation summary
==
> no device associated with this report was received for examination. The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system. Depuy considers the investigation closed. Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
(b)(4). Initial reporter occupation: lawyer. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Pcf and medical records received. After review of medical records patient was revised to addressed mechanical derangement of orthopedic device. Operative notes indicated clicking. There was some metallic wire debris. The base of the neck of the prosthesis of the femur was touching up against the rim of the acetabulum. Had a 30 degrees anteversion of his femur and 25 degrees of his acetabulum but this combination was enough to cause this impingement. Doi: (b)(6) 2007; dor: (b)(6) 2011; right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePINNACLE MTL INS NEUT40IDX58OD
Type of DevicePINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key9660774
MDR Text Key184612389
Report Number1818910-2020-03769
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/31/2012
Device Catalogue Number121887458
Device Lot Number2313450
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/03/2020 Patient Sequence Number: 1
-
-