Since the device was not returned, we are unable to perform further root cause analysis and the exact cause of the reported event is unknown.
All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.
No corrective action.
Monitoring and trending this type of event.
If information is provided in the future, a supplemental report will be issued.
|
Medtronic received report that the distal tip lead wire detached during recapture.
The detached wire was recaptured with the phenom plus and was successfully removed.
Resistance was felt on retrieval.
The patient injury occurred.
The post angiographic results were good.
The devices were prepared and use per the instructions for use (ifu).
The patient underwent treatment of a left internal carotid artery (ica), unruptured, saccular aneurysm.
The distal landing zone was 3.
7 mm and the proximal was 4.
1 mm.
The vessel anatomy was severe in tortuosity.
|