Model Number AU00T0
Device Problem Malposition of Device (2616)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/31/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported during the implant of an intraocular lens (iol) using a preloaded delivery system, the plunger went over the lens.There was contact with the patient and the procedure was completed the same day.Additional information was requested.
 
Manufacturer Narrative
The device with the lens was returned in the opened blister tray inside the opened carton.The plunger lock and lens stop have been removed.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger has advanced the lens to the fill line and has underrode the trailing and right side of the lens.The plunger is advanced toward the right side in the device.The right side of the optic is lifted onto the plunger and the remaining optic extends toward the left side in the device.The trailing haptic is looped around the plunger tip.The leading haptic is misfolded onto the optic and bent forward with the distal haptic tip oriented toward the nozzle tip.The nozzle was cleaned for further evaluation.The lens was removed during cleaning.Top coat dye stain testing was conducted with acceptable results.All product and batch history records are quality reviewed prior to product release.A qualified viscoelastic was used.A plunger underride was observed.This may have been interpreted as the reported complaint.The product investigation could not identify a root cause.A qualified viscoelastic was used.Plunger underride may occur: due to rapid advancement faster than the dfu recommend rate.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to underride the lens.If the device is overfilled with ovd, this can prevent the trailing haptic from being placed properly or move the lens out of position resulting in misfolding.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Top coat dye stain testing was conducted with acceptable results.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key9665040
MDR Text Key181931385
Report Number1119421-2020-00174
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model NumberAU00T0
Device Lot Number12744718
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2020
Date Manufacturer Received03/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNSPECIFIED VISCOAT OVD, 000452, BEL.
-
-