BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U125 |
Device Problems
Failure to Capture (1081); High impedance (1291)
|
Patient Problem
Dizziness (2194)
|
Event Date 09/19/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.
The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.
|
|
Event Description
|
It was reported that this patient experienced a near syncopal event and data review noted non capture.
However, there appears to be dual pacing spikes at a rate of 100ppm and the patient's underlying rhythm is approximately 40bpm.
A boston scientific technical services consultant stated that if there is pacing then there is no capture.
The consultant reviewed the data that was collected today.
There is variable impedance measurements on all channels with no out of range measurements observed.
It appears that signal artifact monitoring (sam) had disabled the respiratory rate trend (rrt); however, there was no stored episode.
It was noted that the patient is in chronic atrial fibrillation (af) but atrial sensing is still programmed on.
The caller was not able to reproduce any issues during in office testing and reprogramming was performed to unipolar pacing and bipolar sensing.
The consultant recommended further evaluation of the system.
No adverse patient effects were reported.
|
|