COOK INC GUNTHER TULIP JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number IGTCFS-65-JUG |
Device Problems
Obstruction of Flow (2423); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Internal Organ Perforation (1987); Sepsis (2067); Thrombosis (2100); Perforation of Vessels (2135); Depression (2361); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
Reporter occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
|
Event Description
|
It is alleged that the patient received a gunther tulip on (b)(6) 2008.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: investigation.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: blood clots, sepsis, depression, anxiety, physical limitations, easily tired and weak breathing.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Unknown if the reported sepsis, depression, anxiety, physical limitations, easily tired and weak breathing are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient further alleges "i'm unable to lift anything over xxxx pounds, am unable to walk and tire out very quickly.I've had several blood clots that have led to sepsis.My breathing is also very weak at times" and "i'm unable to lit weight, hike, kayak or canoe." the patient also alleges depression and anxiety.
|
|
Event Description
|
Patient allegedly received an implant.Patient is alleging tilt, vena cava perforation and abutting of ivc wall.Report from computerized tomography (ct): "ivc filter surveillance." "an infrarenal ivc filter appears intact.Tilting of the filter with the apex abutting the right posterolateral ivc wall.Perforation of 6 to 7mm of the distal aspect of the left lateral strut into the aorta.An anterior strut perforates the ivc wall, and abuts mesenteric vasculature and prominent but nonenlarged aortocaval lymph node.A right lateral strut perforates the ivc wall and abuts the adjacent ascending colon, without appreciable intraluminal bowel perforation." "mild apex right posterolateral tibial of an intact infrarenal ivc filter.A strut perforated the aorta.".
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: a2, b1, b5, b6, d1, d4, h4, h6.H6 device code(s): appropriate term/code not available (3191) was selected for the alleged device perforation.H6 device code(s): appropriate term/code not available (3191) was selected for the alleged device tilt.Additional information: investigation.Investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava (vc)/aorta perforation, tilt, abutting of inferior vena cava (ivc) wall.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported abutting of inferior vena cava (ivc) wall is directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|
|
|