BOSTON SCIENTIFIC CORPORATION EMBLEM S-ICD; IMPLANTABLE LEAD Back to Search Results
Model Number 3501
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581); Failure to Select Signal (1582); Device Sensing Problem (2917)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Event Date 12/14/2018
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that smart pass was off, and this caller requested assistance to program it back on.A boston scientific technical services consultant provided directions and smart pass was re-enabled.Appropriate sensing was confirmed.Further discussion revealed the patient has received some inappropriate shocks due to noise.The consultant recommended further troubleshooting to determine if noise is reproducible and if so, review other sense vectors to see if there is a better option for this patient.No adverse patient effects were reported.
 
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.The 3191 code was utilized due to the surgery that occurred.
 
Event Description
It was reported that this system was explanted and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key9667940
MDR Text Key177762202
Report Number2124215-2020-02594
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526599200
UDI-Public00802526599200
Combination Product (y/n)N
PMA/PMN Number
P110042/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 01/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/08/2020
Device Model Number3501
Device Catalogue Number3501
Device Lot Number113432
Was Device Available for Evaluation? No
Date Manufacturer Received12/04/2020
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age55 YR
-
-