Model Number 3501 |
Device Problems
Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581); Failure to Select Signal (1582); Device Sensing Problem (2917)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
Event Date 12/14/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that smart pass was off, and this caller requested assistance to program it back on.A boston scientific technical services consultant provided directions and smart pass was re-enabled.Appropriate sensing was confirmed.Further discussion revealed the patient has received some inappropriate shocks due to noise.The consultant recommended further troubleshooting to determine if noise is reproducible and if so, review other sense vectors to see if there is a better option for this patient.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.The 3191 code was utilized due to the surgery that occurred.
|
|
Event Description
|
It was reported that this system was explanted and replaced.No additional adverse patient effects were reported.
|
|