Model Number 20
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Patient Involvement (2645)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to physio-control. A third-party service agent evaluated the customer's device and was able to verify the reported issue. However, the repair was rejected by the customer and the device was returned unrepaired. The cause of the reported issue was determined to be the therapy pcb assembly. Device evaluated by the third party agent.
 
Event Description
A third-party service agent contacted physio control to report that their customer's device would deliver monophasic shocks. In this state there will be a partial loss of defibrillator output energy due to a loss of a portion of the biphasic output waveform. The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level. There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of DeviceAUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key9668825
MDR Text Key190780204
Report Number0003015876-2020-00164
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/04/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number20
Device Catalogue Number99402-000017
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/09/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/13/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No