BARD PERIPHERAL VASCULAR, INC. RECOVERY FILTER SYSTEM VENA CAVA FILTER Back to Search Results
Catalog Number RF048F
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Blood Loss (2597)
Event Date 02/11/2017
Event Type  Injury  
Manufacturer Narrative

Manufacturing review: the device history record (dhr) could not be performed as the lot number is unknown. Investigation summary: the device was not returned for evaluation. Medical records were provided and reviewed. Approximately twelve years post filter deployment, ct revealed one of the tines of ivc filter outside the ivc which likely resulted in the large retroperitoneal hematoma. Therefore, the investigation is confirmed for the perforation of the ivc. Based upon the available information, the definitive root cause is unknown. Labeling review: a review of product labeling documents (e. G. Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.

 
Event Description

It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism and in conjunction with or before orthopedic procedure. At some time, post filter deployment, it was alleged that the filter perforating the main artery. The device has not been removed and there were no reported attempts made to retrieve the filter. The patient reportedly experienced internal bleeding; however, the current status of the patient is unknown.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameRECOVERY FILTER SYSTEM
Type of DeviceVENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key9671441
MDR Text Key178357201
Report Number2020394-2020-00850
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK031328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER,HEALTH PROFESSIONAL
Reporter Occupation
Type of Report Initial
Report Date 02/03/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received02/05/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberRF048F
Device LOT NumberUNKNOWN
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/08/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 02/05/2020 Patient Sequence Number: 1
-
-