• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO ENDO PADDLE RETRACT; RETRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO ENDO PADDLE RETRACT; RETRACTOR Back to Search Results
Model Number 173046
Device Problem Difficult to Remove (1528)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, prior to a procedure, the device got stuck in introducer.User opened another device to resolve the issue.There was no patient involvement.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection of the returned product noted that the device had damaged paddle cover.Exposed deformation of the shaft of the introducer was observed.A functional evaluation found that the paddle fully closed and opened by turning of black rotation knob.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of damaged paddle retractor cover may occur when either the paddle retractor is retracted with the paddle in a partially open state; pushing the introducer forward over a paddle that is not fully closed may cause stretching and breakage of the introducer; or if the paddle retractor is subjected to excessive manipulation or used for leverage, these forces may cause the tearing of the cover/ cloth.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO PADDLE RETRACT
Type of Device
RETRACTOR
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key9676302
MDR Text Key177989062
Report Number2647580-2020-00503
Device Sequence Number1
Product Code GAD
UDI-Device Identifier10884523000771
UDI-Public10884523000771
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K914190
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Model Number173046
Device Catalogue Number173046
Device Lot NumberP9A1249Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-