LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/30/2019
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement the heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(6) received a report that a heater-cooler system 3t device was found to be contaminated with mycobacterium chimaera.The lab report confirming the reported contamination has been provided to livanova.The device will be scrapped.There is no known patient involvement.
 
Manufacturer Narrative
H.10: through follow-up communication livanova learned that the device was cleaned as per the instruction for use since when the device was upgraded.It was placed inside the operation theatre during use with the fan directed opposite to the surgery field.The laboratory report confirming the reported contamination by real-time pcr has been provided to livanova.The device was manufactured in 2007 thus no deep disinfection will be conducted to solve the issue because the device will be scrapped.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
MDR Report Key9680294
MDR Text Key178644343
Report Number9611109-2020-00060
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received03/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2076/2081-2015