Model Number TSH
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/24/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).[(b)(6).Pdf].
 
Event Description
The initial reporter stated they received discrepant results for one patient sample tested with the elecsys tsh assay and the elecsys ft4 iii assay on a cobas 8000 e 801 module.The values measured on the e 801 analyzer were reported outside of the laboratory.The tsh value for the sample was unexpected, so it was repeated on an abbott architect analyzer.This medwatch will apply to the tsh assay.Please refer to the medwatch with patient identifier (b)(6) for information related to the ft4 assay.Refer to the attachment for all patient data.The serial number of the customer's e 801 analyzer is (b)(4).
 
Manufacturer Narrative
The investigation could not identify a product problem.The cause of the event could not be determined.Assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and differences in the standardization methodology used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key9681594
MDR Text Key205167614
Report Number1823260-2020-00342
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberTSH
Device Catalogue Number07028091190
Device Lot Number440704
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-