MAUDE Adverse Event Report: KINETIC CONCEPTS, INC. ACTIV.A.C.¿ THERAPY SYSTEM; OMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KINETIC CONCEPTS, INC. ACTIV.A.C.¿ THERAPY SYSTEM; OMP Back to Search Results
Model Number WNDACT
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 01/10/2020
Event Type  Injury  
Manufacturer Narrative
Based on information provided, it cannot be determined that the alleged debridement with increased wound size is related to the activ.A.C.¿ therapy system.Kci has made multiple unsuccessful attempts to obtain additional clinical information.Device labeling, available in print and online, states: precautions if a wound has been progressing well from dressing change to dressing change but then deteriorates rapidly, consider the following interventions and, where necessary, seek the guidance/expertise of a specialist: check the therapy hour meter to ensure that the actual number of therapy hours received matches the number of recommended therapy hours (22 hours a day).If the number of therapy hours is less than 22 each day, find out why there is a therapy deficit and remedy the situation.Clean wound more thoroughly during dressing changes.Evaluate for signs and symptoms of infection and, if present, treat accordingly.Change dressing often, ensuring that it is being changed at least every 48 hours.Examine the wound and debride as necessary.Debride the wound edges if they appear non-viable or rolled under as this may inhibit the formation of granulation tissue and migration of epithelial cells over an acceptable wound base.
 
Event Description
On 10-jan-2020, the following information was reported to kci by the nurse: the patient's wound sized allegedly increased post sharp debridement.The exact date of the debridement was not documented.The patient had no signs or symptoms of infection.The patient continued using v.A.C.® therapy.A device evaluation of the activ.A.C.¿ therapy system is currently pending completion.
 
Manufacturer Narrative
Based on the additional information obtained regarding the device, kci's assessment remains the same: it cannot be determined that the alleged debridement with increased wound size is related to the activ.A.C.¿ therapy system.
 
Event Description
On 10-sep-2019, the device was tested per quality control procedure by kci service center, and the unit passed the quality control checks and met specifications.On (b)(6) 2019, the device was placed with the patient.On 12-feb-2020, the device was tested per quality control procedure by kci quality engineering and the unit passed the quality control checks and met specifications.Inspection and testing of the device did not reveal any evidence of an operational malfunction with the unit that would impact the device's ability to administer v.A.C.® therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIV.A.C.¿ THERAPY SYSTEM
Type of Device
OMP
Manufacturer (Section D)
KINETIC CONCEPTS, INC.
6203 farinon drive
san antonio TX 78249
MDR Report Key9683998
MDR Text Key188480109
Report Number3009897021-2020-00055
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00878237008188
UDI-Public0100878237008188
Combination Product (y/n)N
PMA/PMN Number
K063692
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 03/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWNDACT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2020
Date Manufacturer Received02/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient Weight67
-
-