Based on information provided, it cannot be determined that the alleged debridement with increased wound size is related to the activ.A.C.¿ therapy system.Kci has made multiple unsuccessful attempts to obtain additional clinical information.Device labeling, available in print and online, states: precautions if a wound has been progressing well from dressing change to dressing change but then deteriorates rapidly, consider the following interventions and, where necessary, seek the guidance/expertise of a specialist: check the therapy hour meter to ensure that the actual number of therapy hours received matches the number of recommended therapy hours (22 hours a day).If the number of therapy hours is less than 22 each day, find out why there is a therapy deficit and remedy the situation.Clean wound more thoroughly during dressing changes.Evaluate for signs and symptoms of infection and, if present, treat accordingly.Change dressing often, ensuring that it is being changed at least every 48 hours.Examine the wound and debride as necessary.Debride the wound edges if they appear non-viable or rolled under as this may inhibit the formation of granulation tissue and migration of epithelial cells over an acceptable wound base.
|
On 10-sep-2019, the device was tested per quality control procedure by kci service center, and the unit passed the quality control checks and met specifications.On (b)(6) 2019, the device was placed with the patient.On 12-feb-2020, the device was tested per quality control procedure by kci quality engineering and the unit passed the quality control checks and met specifications.Inspection and testing of the device did not reveal any evidence of an operational malfunction with the unit that would impact the device's ability to administer v.A.C.® therapy.
|