Catalog Number 04625315003
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/16/2020
Event Type  malfunction  
Manufacturer Narrative
The complained test strips have been calibrated against the who standard and are in scope of the roche initiated recall.For these test strips there is a potential for a product problem when the inr is > 4.5.Values > 4.5 inr showed an increasing positive bias.The information in the case is consistent with the details of the recall and the issue has been fully investigated.The related retention test strips were tested in comparison to masterlot #286321-80 on internal reference meters.For this purpose, two human blood samples from marcumar donors and two internal reference meters were used.No error messages occurred.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." this event occurred in (b)(6).Occupation was lay user/patient.
 
Event Description
The initial reporter received questionable results from coaguchek xs meter serial number (b)(4).At 12:26 pm, the meter result was 7.7 inr.At 1:30 pm, the result from a laboratory using an unknown method was 5.7 inr.The therapeutic range was 2.0 - 3.0 inr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key9687608
MDR Text Key219783608
Report Number1823260-2020-00372
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/31/2020
Device Catalogue Number04625315003
Device Lot Number33046111
Was Device Available for Evaluation? No
Date Manufacturer Received01/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberRES Z-0360-2019
Patient Sequence Number1
Patient Age79 YR
Patient Weight65
-
-