MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX EMBOLIZATION DEVICE INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX EMBOLIZATION DEVICE INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED2-450-16
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/07/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that after the pipeline was implanted and the positioning was confirmed, it was attempted to recapture the carrier guide. The guide on which the pipeline was mounted was detached from the distal delivery wire. It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event. The patient was undergoing surgery for treatment of an unruptured aneurysm in the left internal carotid artery with a max diameter of 4. 3mm, a 3. 2mm neck diameter, and a 3. 76x4. 5mm landing zone. Ancillary devices include a xt27 microcatheter, navien microcatheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePIPELINE FLEX EMBOLIZATION DEVICE
Type of DeviceINTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
katcha taylor
9775 toledo way
irvine, CA 92618
9496801345
MDR Report Key9688925
MDR Text Key189191683
Report Number2029214-2020-00130
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/15/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberPED2-450-16
Device Catalogue NumberPED2-450-16
Device Lot NumberA862902
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-