• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3210
Device Problems Low impedance (2285); Capturing Problem (2891); Appropriate Term/Code Not Available (3191); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/22/2020
Event Type  malfunction  
Manufacturer Narrative
The pacemaker cap confirm was automatically disabled with no abnormal right ventricular lead impedance.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented to the clinic for routine pacemaker check.Upon interrogation of the device, it was discovered that a polarity switch occurred on the left ventricular lead and the cap confirm setting was automatically disabled on the left and right ventricular leads.It was noted that the left ventricular lead exhibited low lead impedance causing the polarity switch.The right ventricular lead, however, did not exhibit any abnormal lead impedance.The pacemaker output also increased for both lead chambers.The leads were manually tested and found to have stable impedance and threshold.It was noted that the patient underwent some radiation therapy during the time of the event.The device and lead were then reprogrammed back to the appropriate settings.There were no adverse consequences to the patient.Related manufacturer reference number: 2017865-2020-01823.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9690860
MDR Text Key178574414
Report Number2017865-2020-01822
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734503785
UDI-Public05414734503785
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Model NumberPM3210
Device Catalogue NumberPM3210
Device Lot Number4397522
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/22/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/19/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUICKSITE
Patient Age74 YR
-
-