Concomitant medical products: erbe electrosurgical unit, vio 200s.Investigation evaluation: our laboratory evaluation of the product said to be involved could not confirm the report as it was described.The handle of the device was manipulated and the snare head would open and close with no resistance felt.During a visual inspection of the device, there was no damage noted on the device.The continuity from the electrical pin to the snare head was tested with an ohm meter and passed.An additional functional test was performed by attaching the active cord to the electrical pin.The active cord connected to the device easily and remained securely connected.The device was connected to a valley lab generator and power was applied.The snare cut simulated tissue as expected.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for the reported observation could not be determined because the product said to be involved functioned as intended when tested with simulated tissue.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.The instructions for use contain the following information to assist with proper set-up and use of the device: "before using this device, follow recommendations provided by electrosurgical unit manufacturer to ensure patient safety through proper placement and utilization of patient return electrode.Ensure a proper path from patient return electrode to electrosurgical unit is maintained throughout the procedure." "inspect active cord.Cord must be free of kinks, bends, breaks and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." "with electrosurgical unit off, prepare equipment.Securely connect active cord to device handle and electrosurgical unit.Active cord fittings should fit snugly into both device handle and electrosurgical unit.Following instructions from electrosurgical unit manufacturer, position patient return electrode and connect it to electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify desired settings and activate electrosurgical unit.Note: maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
During a colonoscopy, the physician used a cook acusnare polypectomy snare.While snaring a large polyp, the physician had to lift it with ellevue.Then when the polyp was snared off, the physician then had to cauterize where the polyp was snared.The device began to spark.A new snare was used to complete the procedure.A clip was placed after the polypectomy.The body pad was on the patient's outer thigh.The patient's position changed from left lateral to supine and they did not readjust the pad but the light was green and there was connectivity.The cook district manager (dm) observed and there appeared to be tissue inside the snare at the tip and believes this is what caused the spark.A section of the device did not remain inside the patient¿s body.The patient required a clip after the polypectomy to stop the bleeding.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|