Model Number 4FC12 |
Device Problems
Fluid/Blood Leak (1250); Gas/Air Leak (2946)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, air ingress and leaking occurred.The sheath was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the sheath 4fc12 with lot number 70896 was returned and analyzed.Visual inspection showed the device was intact with no apparent issue.Air aspiration was reproduced during the aspiration test when the test catheter was introduced through the sheath.The hemostatic valve was leaking; and it was suspected the valve disk is torn.Further dissection didn't show any leak along the shaft in the handle.In conclusion, the reported air ingress issue was confirmed through testing.The sheath failed the returned product inspection due to a leaking hemostatic valve.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Correction: h6.If information is provided in the future, a supplemental report will be issued.
|
|