• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAZOR ROBOTICS LTD RENAISSANCE SYSTEM; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAZOR ROBOTICS LTD RENAISSANCE SYSTEM; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number TPL0038
Device Problem Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/06/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure.It was reported that the case was for two vertebrae on level l4 - l5.There were no issues with attachment.During registration, new images were taken in order to get registration.L5 was registered on first registration but l4 was greater than 1.3.L5 left was superior and plan was adjusted accordingly at l5 left.L4 left was also superior.The whole plan was adjusted accordingly.All neuromonitoring values were over 20.No patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENAISSANCE SYSTEM
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IL  3079567
Manufacturer (Section G)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IL   3079567
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9712094
MDR Text Key204720733
Report Number3005075696-2020-00032
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120812
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTPL0038
Device Catalogue NumberTPL0038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/06/2018
Date Device Manufactured01/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age78 YR
-
-