Model Number TPL0038
Device Problem Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/23/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure.It was reported that the case was planned prior to surgery for two vertebrae on levels l4 - l5.Attention was paid to axial angles and skin incisions.The bed mount was chosen for this case due to clinical indication and surgeon preference.Auto segmentation was achieved using basic algorithm acquiring good values at l4 but a value greater than 1.3 at l5.The representative went back and made manual adjustments to segmentation lines.Great values were then acquired at both levels.All trajectories were sent, and wires were placed.Images were taken and both l4 wires were very low.The surgeon continued to place l5 screws and free handed both l4 screws.Surgeon was pleased with final images.No patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENAISSANCE SYSTEM
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IL  3079567
Manufacturer (Section G)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IL   3079567
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9712556
MDR Text Key192177665
Report Number3005075696-2020-00057
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120812
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTPL0038
Device Catalogue NumberTPL0038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2018
Date Device Manufactured07/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age41 YR
Patient Weight104
-
-