COVIDIEN MFG SOLUTIONS S.A. MAHURKAR CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number 8830415001
Device Problems Material Frayed (1262); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during catheter insertion, the guide wire was noted to be branching and unraveled and cannot be pulled out. The catheter was flushed prior to use and the lines were not reversed. The dimensions of the catheter was stated to match what was indicated in the label and the guide wire provided in the kit was the one used during the insertion. The catheter was not repaired and there was no leakage. It was also stated that excessive force was required to remove the guide wire, and the guide wire and the catheter were removed from the patient simultaneously (in one action). The guide wire was completely removed and locked in the needle and was no longer intact. Post-operative x-ray was not performed. There were no patient symptoms associated with the event. The catheter was replaced in order to resolve the issue. There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMAHURKAR
Type of DeviceCATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
Manufacturer Contact
lisa hernandez
15 hampshire street
mansfield, MA 02048
2034925563
MDR Report Key9720370
MDR Text Key193400616
Report Number3009211636-2020-00051
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K943349
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number8830415001
Device Catalogue Number8830415001
Device Lot Number1703200096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/13/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-