Catalog Number ASK-05500-KM |
Device Problem
Difficult to Flush (1251)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qn#: (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
|
Event Description
|
It was reported that the user was unable to flow saline into the catheter with syringe during a flushing test.Therefore, it was replaced with a new kit.
|
|
Event Description
|
It was reported that the user was unable to flow saline into the catheter with syringe during a flushing test.Therefore, it was replaced with a new kit.
|
|
Manufacturer Narrative
|
Qn#(b)(4).A device history record review was performed on the epidural catheter with no relevant findings.The customer reported the catheter would not flush.The customer returned 5ml one syringe, one snaplock assembly, an epidural catheter, and lidstock.The returned snaplock assembly and catheter were received connected together (reference files (b)(4).The returned components were visually examined with and without magnification.Visual examination of the returned snaplock assembly revealed the snaplock appears typical with no observed defects or anomalies.Visual examination of the returned catheter revealed the catheter appears used.Biological material can be seen between the inner coils.A functional flow test was performed on the returned sample per amrq-000017 section 7.8; rev.7.The returned epidural catheter was inserted from the proximal end into the returned snaplock assembly until it bottomed out and the snaplock assembly was closed.The components were confirmed to be secured by tugging gently.The snaplock assembly was connected to the lab leak tester (ref-002902) and the pressure was increased to 10 psi to establish flow.Water could be seen immediately exiting the distal end of the catheter.The flow rate was measured at 9.0ml/min (c05180), which is within the specification of 1ml/min.No blockages were found.A corrective action is not required at this time as there were no functional issues found with the returned sample.The reported complaint of the catheter not flushing could not be confirmed based on the sample received.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related issue.The returned components passed a functional flow test and met flow rate specifications.There were no functional issues found with the returned sample.
|
|
Search Alerts/Recalls
|