RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 18feb2020.
|
|
Event Description
|
It was reported that the ventilator had an alarm for blower high temperature.There was no patient involvement.
|
|
Manufacturer Narrative
|
G4: 23jun2020.B4: 24jun2020.A blower was return for analysis.A visual inspection of the returned component was performed, no notable conditions were found.An investigation was performed and the product analysis technician reported that no fault was found.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4: 02mar2020.B4: 05mar2020.The manufacturer¿s international service technician confirmed the reported issue.The service technician found the vibration dampening ring was not functioning as intended and therefore, the blower temperature was raised by the resonating vibration occurring at the time of noise generation.The manufacturer¿s international service technician replaced the blower and the vibration dampening ring was adjusted to address the reported problem.The unit was checked overall, run in tested, cleaned, functionally tested and no abnormality was confirmed.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|