The device has been received; however, the analysis has not been completed.
The cause of the reported event has not been determined at this time.
Once the analysis has been completed, a supplemental report will be submitted.
If information is provided in the future, a supplemental report will be issued.
|
Medtronic received report that the distal tip of the delivery wire of the medtronic flow diverter separated, and detached at the solder joint post successful deployment.
The devices were prepared and use per the instructions for use (ifu).
The angiographic result post the procedure showed the ped-375-20 correctly and successfully placed over medium sized wide-necked aneurysm in the left internal carotid artery at the terminus segment.
The patient was doing well post procedure.
The medtronic catheter and delivery wire with detached tip of the flow diverter device will be returned for analysis.
The patient underwent treatment of a left internal carotid artery (ica), at the level of the terminus segment of the ica, unruptured, saccular aneurysm.
The max diameter was 4 mm x 4 mm and the neck was 3 mm.
The distal landing zone was 4 mm and the proximal was 5 mm.
The vessel anatomy was minimal in tortuosity.
|