Model Number EPIQ 5G - 795204
Device Problem Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An investigation is underway to determine the root cause of the issue.Results of the investigation will be included in a follow up report upon its completion.Data from the customer is being collected and analyzed.
 
Event Description
A customer reported encountering an incident where a patient¿s data was mixed with another exam.According to the feedback details, an image from a previous study appeared in the next patient¿s exam.The data mix-up was identified by the user and there was no allegation of altered patient outcome as a result of the issue.
 
Manufacturer Narrative
A thorough investigation performed by development engineering identified the reported issue as a software anomaly caused by a race condition in the capture code between thumbnail and still frame capture operations.A solution for this issue has been included in a newly released software revision.The site¿s local philips service team has been notified through normal service channels.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.H3 other text : data from the customer has been collected and analyzed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ 5G - 795204
Type of Device
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC22
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
MDR Report Key9729681
MDR Text Key184130138
Report Number3019216-2020-00010
Device Sequence Number1
Product Code IYN
Combination Product (y/n)N
PMA/PMN Number
K160807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 01/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEPIQ 5G - 795204
Device Catalogue NumberEPIQ 5G - 795204
Device Lot NumberUS315C0721
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-