WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-FEM-TULIP
Device Problems Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); Pain (1994); Perforation of Vessels (2135); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Manufacturers ref# b)(4).Catalog# is unknown but referred to as cook gunther tulip filter.Occupation: non-healthcare professional.Pma/510(k): k172557.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook gunther tulip filter on (b)(6) 2017.It is alleged that the [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.".
 
Event Description
Patient allegedly received an implant via right common femoral vein due to deep vein thrombosis (dvt).Patient is alleging tilt, vena cava perforation, embedment.The patient further alleges back pain and stress.Per a computerized tomography (ct) scan of the abdomen and pelvis dated (b)(6) 2018, ¿the filter is tilted anteriorly.Its proximal cone lies on the wall of the ivc possibly embedded in it.The filter legs have penetrated through the wall of the ivc into the pericaval/mesenteric fat.¿.
 
Manufacturer Narrative
Device code(s): appropriate term/code not available (3191) was selected for the alleged device tilt.Device code(s): appropriate term/code not available (3191) was selected for the alleged device perforation.Investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava (vc) perforation, embedment, tilt, back pain, stress.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported back pain and stress are directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.No other complaints on lots.Product is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key9731643
MDR Text Key183138905
Report Number3002808486-2020-00227
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K090140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 03/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/27/2019
Device Catalogue NumberIGTCFS-65-1-FEM-TULIP
Device Lot NumberE3499278
Was Device Available for Evaluation? No
Date Manufacturer Received02/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight145
-
-