Reportable based on device analysis completed on 30-jan-2020.It was reported that tip damage occurred.The 85% stenosed, 18mmx2.5mm target lesion was located in the severely tortuous and moderately calcified distal end of left anterior descending artery.A 2.00mm x 15mm emerge balloon catheter was selected for use.However, during procedure, the tip of the balloon got kinked.The procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient's status was stable.However, returned device analysis revealed balloon pinhole.
|