• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ALTRX +4 NEUT 36IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ALTRX +4 NEUT 36IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1221-36-452
Device Problem Difficult to Insert (1316)
Patient Problem No Code Available (3191)
Event Date 02/05/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Surgeon was unsuccessful getting 52x36 +4n altrx liner to seat fully on several attempts.There was originally a hole in eliminator and he removed that and still could not get the liner to seat.We decided to remove the liner and opened a new liner and it seated correctly on the first try.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: visual examination of the returned device finds nothing outward to suggest product error.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX +4 NEUT 36IDX52OD
Type of Device
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9733956
MDR Text Key190343192
Report Number1818910-2020-05487
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295016489
UDI-Public10603295016489
Combination Product (y/n)N
PMA/PMN Number
K062148
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1221-36-452
Device Catalogue Number122136452
Device Lot NumberJ62P22
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2020
Date Manufacturer Received03/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
APEX HOLE ELIM POSITIVE STOP
Patient Age73 YR
-
-